Standard Test Method for Transition Temperatures and Enthalpies of Fusion and Crystallization of Polymers by Differential Scanning Calorimetry
STANDARD published on 1.5.2015
Designation standards: ASTM D3418-15
Note: WITHDRAWN
Publication date standards: 1.5.2015
SKU: NS-590617
The number of pages: 7
Approximate weight : 21 g (0.05 lbs)
Country: American technical standard
Category: Technical standards ASTM
Keywords:
crystallization, crystallization temperature, differential scanning calorimetry (DSC), enthalpy, first-order transition, glass transition, heat of crystallization, heat of fusion, heat of transition, melting, melting temperature, polymer, transition temperature,, ICS Number Code 83.080.01 (Plastics in general)
Significance and Use | ||||||||||||||||||
5.1 Thermal analysis provides a rapid method for measuring transitions due to morphological or chemical changes, in a polymer as it is heated/cooled through a specified temperature range. Change in specific heat capacity, heat flow and temperature values are determined for these transitions. Differential scanning calorimetry is used to assist in identifying specific polymers, polymer alloys, and certain polymer additives, which exhibit thermal transitions. Chemical reactions that cause or affect certain transitions have been measured with the aid of this technique; such reactions include oxidation, curing of thermosetting resins, and thermal decomposition. 5.2 This test method is useful for specification acceptance, process control, and research. |
||||||||||||||||||
1. Scope | ||||||||||||||||||
1.1 This test method covers determination of transition temperatures and enthalpies of fusion and crystallization of polymers by differential scanning calorimetry. Note 1: True heats of fusion are to be determined in
conjunction with structure investigation, and frequently,
specialized crystallization techniques are needed.
1.2 This test method is applicable to polymers in granular form or to any fabricated shape from which it is possible to cut appropriate specimens. 1.3 The normal operating temperature range is from the cryogenic region to 600°C. Certain equipment allows the temperature range to be extended. 1.4 The values stated in SI units are the standard. Note 2: This test method does not apply to all types of
polymers as written (see 6.8).
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Note 3: This standard is similar but not equivalent to
ISO 11357-1, -2, -3. The ISO procedures provide additional
information not supplied by this test method.
|
||||||||||||||||||
2. Referenced Documents | ||||||||||||||||||
|
Do you want to make sure you use only the valid technical standards?
We can offer you a solution which will provide you a monthly overview concerning the updating of standards which you use.
Would you like to know more? Look at this page.
Latest update: 2024-04-17 (Number of items: 2 859 866)
© Copyright 2024 NORMSERVIS s.r.o.